- Advantage : You can simply make a 3D implant guide customized for your patient in the middle of your surgery.
- You can also select and assemble a guide respectively, depending on the crown diameter even in a multi case.
- You can plan the crown size and prosthetic appliance at the same time of implant placing.
- You can solve the depth and distance interval with only one drilling.
- Lego system in consideration of the prosthetic size.
- Ideal Stopper Design & Crown Guide Design.
Direction for use
- Trephine bur : Ø4.5, Ø5.0, Ø5.5, Ø6.0 – Used after installation to the engine handpiece when removing the implant from the mouth.
- Claw Driver : Used by turning counterclockwise to the broken surface of the abutment screw.
- Screw Drill : Drilling (1500~2000RPM) counterclockwise through the screw drill where the screw guide is installed when the broken not removed even with the claw driver.
- Tap Driver : Used by turning counterclockwise after drilling to the broker abutment screw.
- 0.7 Slot Driver : Used by turning counterclockwise to the broken abutment screw surface after drilling.
- Screw Guide : It aims at protecting the thread in the implant with the fixing to the implant.
type : Active RP, External 2.7 Hex, 8° Morse Taper, Submerged 2.5 Hex, TSV 3.7/4.1, Camlog Tube-in-Tube - Implant Remover : 3EA – Used for removing the implant by connecting to the ratchet, putting in the implant and turning counterclockwise.
- Hand / Strauman Adapter : Used for switching the driver for the handpiece to the hand or connection for the use of Strauman ratchet wrench.
- Driver Hand Adapter : Used for switching the driver for the handpiece to the hand or connection for the use of Strauman ratchet wrench.
- Implant Forcep : Used first for attempting the removal of an implant as the exclusive forceps for the implant removal.
- Guide Wrench : Used for fixing the screw guide to the implant.
Related Product
- 10000Direction for use COMPONENT Trephine : 3.8/4.5, 4.3/5.0, 4.8/5.5, 5.3/6.0 Used with connection to the engine handpiece when the implant is removed from the mouth. Implant Remover : Used for manual removal by turning conterclockwise with connection to the ratchet. [Small, Medium, Large] Hand / Strauman Adapter : Used for fixing the remover with hands in the implant or for connecting when Strauman Ratchet Wrench is used. Implant Forcep : Forceps dedicated to the implant used for removing the implant from the mouth.
- 10000Direction for use COMPONENTClaw Driver : Used by turning counterclockwise to the broken surface of the abutment screw.Screw Drill : Drilling (1500~2000RPM) counterclockwise through the screw drill where the screw guide is installed when the broken screw is not removed even with the claw driver.Tap Driver : Used by turning counterclockwise after drilling to the broker abutment screw.Screw Guide : It aims at protecting the thread in the implant with the fixing to the implant. Type : 8° Morse Taper, Submerged 2.5 Hex, TSV 3.7/4.1, Replace RPGuide Wrench : Used for fixing the screw guide to the implant. Used in combination as it is seperately installed to the bottom of the mini kit.